Quantitation of Eight Anticoagulant Rodenticides in Liver

Megan Romano

Published: 2024-01-25 DOI: 10.17504/protocols.io.14egn26w6g5d/v1

Disclaimer

Reference to any commercial materials, equipment, or process does not in any way constitute approval, endorsement, or recommendation by the Food and Drug Administration.

Abstract

Target analytes comprised two chemical classes: hydroxycoumarins (warfarin, coumachlor, dicoumarol, bromadiolone, brodifacoum, and difethialone) and indanediones (diphacinone and chlorophacinone).Liver extracts were cleaned using dispersive solid phase extraction (d-SPE) to remove matrix interferences and analyzed by reverse phase ultraperformance liquid chromatography−tandem mass spectrometry (UPLC−MS/MS). Electrospray ionization in negative ion mode combined with multiple reaction monitoring (MRM) using a triple quadrupole mass spectrometer provided simultaneous confirmation and quantitation. Detection limits spanned 0.75−25 ng/g, and lower quantitation limits were established as 50 ng/g.

Method validation/evaluation/verification:

In-house method validation data and evaluation by an independent laboratory (Vet-LIRN) are published:

https://pubs.acs.org/doi/10.1021/acs.jafc.7b02280

Attachments

Steps

Prepare Solutions

1.

Methanol in Acetonitrile 10% (v/v) Methanol in Acetonitrile: Dilute 25mL methanol to 250mL with acetonitrile.

2.

Primary Stock Solutions – 1000 ug/mL : For each anticoagulant rodenticide, dissolve 5.0mg standard reference material in 5mL of the appropriate solvent (see Table 1), using a 5-mL volumetric flask. These eight solutions should be stored at -20°C and can be stored for up to one year.

Table 1: Solvent Selection Guide

AB
Anticoagulant RodenticideSolvent
Bromadiolone, Coumachlor, WarfarinMethanol
Brodifacoum, Chlorophacinone, Difethialone, DiphacinoneAcetone
DicoumarolChloroform
3.

Secondary Stock Solution – 10 ug/mL : Transfer 50µL of each primary stock solution to a single 5-mL volumetric flask. Complete the volume with methanol to prepare a single solution that is 10 ug/mL of each ACR. This solution should be stored at -20°C and can be stored for up to one month.

Note
The use of a positive displacement pipette may be necessary for accurate transfer of acetone- and chloroform-containing solutions

4.

Mobile Phase Solutions : De-gas mobile phase solutions by helium sparging as needed.

4.1.

a. 0.01M Ammonium Acetate, pH 9 – Dissolve 0.77g ammonium acetate in ~750ml distilled, deionized water (DDI water) in a 1-L volumetric flask. Adjust pH to 9 by adding ammonium hydroxide dropwise. Complete volume with DDI water. This solution may be stored at room temperature for up to one month.

4.2.

b. Methanol, HPLC grade - This solution may be stored at room temperature for up to 5 years.

Prepare Matrix-Matched Calibrants and Quality Control (QC) Samples

5.

Transfer 1.0g pre-homogenized control liver to a 50-mL disposable centrifuge tube compatible with the Geno/Grinder. Repeat this step for a total of 7 calibrants and 3 QC samples.

6.

Pipette out the appropriate volume of Secondary Stock Solution, as described in Table 2, into the corresponding tube and directly into the homogenized liver.

Table 2: Calibrant and Control Sample Preparation Guide

ABC
Calibrant / QC SampleConcentration (ppb; ng/g)Volume of Secondary Stock Solution (uL)
Cal 1252.5
Cal 2505
Cal 3757.5
Cal 410010
Cal 550050
Cal 61000100
Cal 72500250
QC Blank00
QC 50 ppb505
QC 1000 ppb1000100
7.

Cap tubes and vortex mix the spiked liver for 0h 0m 10s to mix thoroughly.

8.

Proceed to step 10

Sample Homogenization and Extraction

9.

Transfer 1.0g unknown sample liver to a 50-mL disposable centrifuge tube.

10.

Homogenize the calibrants, QC samples, and unknown samples as follows:

10.1.

Add a 9.5mm steel ball bearing to each tube and cap.

10.2.

Grind tissue at a rate of 650 rpm for 0h 5m 0s. One round of impact

grinding may be sufficient, however a second round of grinding may be performed

if needed.

11.

Add 6mL 10% (v/v) methanol in acetonitrile to each tube and vortex mix for 0h 0m 10s to mix thoroughly.

12.

Repeat the homogenization process as described in step 10.2.

A single repetition of impact grinding at 650 rpm for 0h 5m 0s is usually sufficient.

13.

Load all calibrants, QC samples, and unknown samples on the reciprocating shaker for 0h 30m 0s.

14.

Centrifuge at 829g (2000 RPM using the specified centrifuge) for 0h 5m 0s

Sample Clean-up with d-SPE

15.

Transfer the supernatant of each extraction mixture to the corresponding d-SPE tube by pouring carefully and recap the d-SPE tube.

16.

Vortex for 0h 0m 10s to wet the sorbent material completely.

17.

Load all calibrants, QC samples, and unknown samples on the multi-mixer & rotator for 0h 30m 0s.

18.

Centrifuge at 829g (2000 RPM using the specified centrifuge) for 0h 5m 0s.

Solvent Exchange / Concentrating Extracts

19.

Transfer the supernatant of the d-SPE clean-up mixture to a 15-mL disposable centrifuge tube.

Note
The supernatant may be carefully poured from the d-SPE tube to the clean tube.

20.

Evaporate to dryness under a gentle stream of nitrogen at 45°C.

Note
Near the end of the drying process, the solution may foam. As long as no bulk solution exists and foam is of negligible volume, the sample preparation process can continue.

21.

Transfer 1.0mL methanol to each tube and vortex for 0h 0m 10s. Sonicate in a water bath for no more than 0h 5m 0sto completely dissolve the residue. Required time for sonication may vary based on unique sample characteristics.

22.

Transfer each prepared solution to a clean disposable microbeaker.

23.

Draw up the entire volume into a clean 5-mL Luer-lok syringe.

24.

Filter the sample through a 0.22 um PTFE syringe filter into an autosampler vial for analysis.

UHPLC – MS/MS Analysis

25.

UHPLC Settings

Gradient Elution Profile : Profile parameters may be adjusted slightly at the discretion of the analyst to achieve baseline resolution of brodifacoum and difethialone at 2500ppb (Cal 7). The suggested gradient profile is in Table 3.

Table 3: Gradient elution profile

ABC
Time (min)0.01M Ammonium Acetate, pH 9 (%)Methanol (%)
06040
16040
94357
152377
181981
191090
241090
256040
346040
25.1.

Flow Rate: 0.300mL/min

25.2.

Column Temperature: 25°C

25.3.

Injection Volume: 1µL

25.4.

Total Run Time: 0h 34m 0s

26.

MS/MS Detection

These parameters are suggested settings and may need to be optimized for different MS instruments

26.1.

ESI Source Conditions :Optimized on the basis of direct infusion of solvent-diluted reference standards

a. Negative ion mode

b. Spray Voltage: 4000V

c. Vaporizer Temperature: 380°C

d. Sheath Gas Pressure: 50 psi

e. Auxillary Gas Pressure: 45 psi

f. Ion Sweep Gas Pressure: 0 psi

g. Capillary Temperature: 300°C

h. Skimmer Offset: (Not used)

26.2.

Other Parameters

a. Collision Gas Pressure: 1.7 mTorr

b. Collision Energy: Ion-Dependent; see MRM Transitions Table (Table 4)

c. Tube Lens: Ion-Dependent; see MRM Transitions Table (Table 4)

d. Q1 / Q3 Peak Width (FWHM): 0.70 u

e. Cycle Time: 0.300 s

Table 4. MRM Monitored Transitions / Expected Retention Times

ABCDEF
Anticoagulant RodenticideRetention Time (min)Precursor Ion ((M-H+)-; u)Fragment Ion (u)Collision Energy (eV)Tube Lens (V) Tube Lens (V)
Warfarin2.00307161*2270
3072502570
Coumachlor4.873412842671
3411612371
Diphacinone7.483391672877
3391654877
Dicoumarol7.683351612147
3351174747
Chlorophacinone11.083732012476
3731452576
Bromadiolone**13.765252503897
5252734097
Brodifacoum16.2852113540101
52114357101
Difethialone16.5353715141100
53737135100

*Transitions in bold are used for quantitation**Two isomers are present for Bromadiolone; only the earliest eluting (and most abundant) isomer is used for detection and quantitation

27.

Post-Acquisition Data Analysis

Note
Automated peak integration should be used as much as possible to minimizemanual data integration.  Automated integration parameters should be determined based upon those that provide accurate peak integration in sample and calibrant chromatograms and may vary between analytes.

27.1.

Qualitative Identification :  Comparisons of unknown chromatograms to those of calibrants and QC samples are used to determine the presence or absence of each analyte.  For an analyte to be considered present in a sample, the following requirements must be met:

a. Retention Time :

  1. The retention times of the primary (Quan) and secondary (Confirming) ions are within 0.25 min of the mean retention time for the corresponding analyte in all calibrants and overspiked QC samples acquired within the same batch.
  2. The primary (Quan) ion and the corresponding secondary (Confirming) ion co-elute within 0.1 min of one another.

b. Signal Intensity :

  1. The peak areas in the primary and secondary ion channels must have a S/N ratio of 3 or greater.
  2. If there are detectable AR levels in any of the associated blank samples (e.g. instrument blanks, reagent blanks, and negative control samples), then the peak area in the unknown sample must be at least 10 times that measured in the blank sample.

c. Ion Ratio Measurement : The ion ratios of the primary to secondary ions for a given analyte, expressed as a percentage, must be within 20% of the average ion ratio for the associated spiked matrix-matched samples (e.g. calibrants, QC samples) acquired concurrently.

27.2.

Quantitative Analysis: Recommended method

Perform linear least squares regression using peak areas for all calibrants versus concentration, ranging from 25 to 2500 ppb. The following parameters should be used to generate calibration curves to determine quantitative results:

a. Weighting 1/x2

b. Ignore Origin

c. The coefficients of determination (R2) for each calibration are expected to be ≥0.985 when the fit is linear and the origin (0,0) is ignored.

Note
Occasionally, a quadratic instrument response is observed at concentrations > 1000 ppb. It is permissible to use a quadratic trendline, with or without weighting, in these situations with an R2 value of 0.985 or better.

d. The intensity of the quantifying ion is greater than the intensity of the same ion in the least concentrated calibrant.

Note
Overspiked QC samples (e.g., 50 ppb and 1,000 ppb) must yield AR concentration results that are within acceptable range (e.g., +/- 15%) of the expected AR concentrations

推荐阅读

Nature Protocols
Protocols IO
Current Protocols
扫码咨询