Air Compressor Use and Maintenance

Ksenija Sabic

Published: 2023-12-27 DOI: 10.17504/protocols.io.81wgbxwm3lpk/v1

Abstract

This manual includes instructions for regular use and maintenance of ACGP Series Air Compressor by Newport.

Instructions are taken directly from manufacturer website: https://www.newport.com/p/ACGP (also attached).

Please use the printed Air Compressor Maintenance Checklist in Alcove D to record maintenance activity.

Attachments

Steps

Operation (before and after use of Xenium instrument)

1.

Turn on (if not already)

Make sure that the power supply voltage is the same with what’s indicated on the data label: 230V/50 Hz or 115V/60 Hz. Use only 3-pole receptacles that accept the compressor’s plug.

The on/off switch is located on the top of the pressure switch. To switch the unit on, turn the switch to position “1”. When switched on, the compressor starts to build up pressure until the tank reaches the maximum allowed pressure (8 bar/120 psi). The unit will then be automatically

switched off until air is used and the pressure in the tank drops to the minimum operating pressure (6 bar/90 psi), at which point the compressor automatically switches on.

2.

Check the outlet pressure

The outlet pressure can be adjusted using the knob on the filter/regulator so that the air output is enough for the application, but the output will not exceed the maximum air pressure specified by the application. Turn the knob clockwise to increase the pressure and counter-clockwise to decrease pressure.

Weekly Maintenance

3.

Check for oil leaks, abnormal noise or vibration.

3.1.

Check oil level through the sight glass weekly to make sure the oil level is aligned with the red marker line next to the sight glass. Add or remove oil as needed.

Monthly Maintenance

4.

Check oil level.

Drain the condensate that has collected in the air tank. It may be necessary to do more often when operated in high humidity.

Drain the water collected in the air outlet filter.

Clean the compressor using a soft cloth. Wipe off any oil or dirt.

Check compressor efficiency (make sure all nuts and bolts are tight, check for air leaks, pressure hose wear, pipe fittings and electrical connections).

Quarterly Maintenance

5.

Check the air intake filter. Replace if necessary.

Annual Maintenance

6.

Replace oil.

Check pressure relief valve function.

推荐阅读

Nature Protocols
Protocols IO
Current Protocols
扫码咨询