Proteolytic Peptide Desalting with C18 Hydrophilic–Lipophilic Balance (HLB) Cartridges
J P Rose, M A Watson, B Schilling, Joanna Bons
Abstract
Desalting of proteolytic peptide elution using C18 Hydrophilic–Lipophilic Balance (HLB) Cartridges in preparation for downstream proteomic profiling.
Steps
Centrifuge the samples at 1,850 x g for 5 minutes at room temperature to pellet insoluble material.
Desalt the samples using Oasis HLB solid-phase extraction cartridges placed on top of a vacuum manifold as follows:
Condition each cartridge two times with 800 µL of the condition buffer (50% ACN, 0.2% FA).
Equilibrate each cartridge three times with 800 µL of the equilibration buffer (0.2% FA).
Load the peptide samples.
Wash each cartridge three times with 800 µL of the wash buffer (0.2% FA).
Elute peptides with 800 µL of the elution buffer (50% ACN, 0.2% FA) followed by a further 400 µL of the same elution buffer.
Vacuum dry the eluted peptide solution in a centrifugal vacuum concentrator.
Reconstitute the dried peptides in 0.2% FA for a final protein concentration of 1 µg/µL and thoroughly mix the solution.
Centrifuge at 12,000 x g at room temperature for min, and store at -20°C.