Large volume viral RNA extraction using MagMAX Viral RNA Isolation Kit

Joyce Akello, Javier Martin, Nick Grassly, Erika Bujaki, Dimitra Klapsa, Thomas Wilton, Alison Tedcastle

Published: 2024-03-21 DOI: 10.17504/protocols.io.81wgbzdr3gpk/v1

Abstract

This method describes large volume nucleic acid purification from sewage concentrates starting with 1.2 mL of sample using MagMAX TM Viral RNA Isolation Kit. To perform automated extraction, follow steps in Workflow A, and for manual extractions follow Workflow B diverging after the initial sample processing with Lysis/Binding solution. Appendix 1 contains pictures providing a visual guide to manual magnetic beads based nucleic acid extraction.

Steps

1.

Reagent

preparation

1.1.

Wash Solution 1

  • Add indicated volume of 100% Isopropanol to the bottle of Wash Solution 1 Concentrate.
  • Mix well by inverting at least 5 times and mark bottle to indicate that the alcohol was added.
1.2.

Wash Solution 2

  • Add indicated volume of 100% ethanol to the bottle of Wash Solution 2 Concentrate.
  • Mix well by inverting at least 5 times and mark the bottle to indicate that ethanol was added.

Note
Taking an aliquot of the prepared Wash Solutions is recommended to avoid potential contaminationif they are to be used on multiple occasions.

1.3.

Lysis/Binding Solution

Combine the components listed below in the order indicated.

Prepare enough solution for the number of samples extracted that day, including controls, with allowing extra for pipetting loss.

  • Add Carrier RNA to Lysis/Binding Solution Concentrate according to the table below, and mix briefly.
AB
ReagentVolume per sample
Lysis/Binding Soln. Concentrate1.2 ml
Carrier RNA6 μl
  • Add 100% Isopropanol and mix well by vortexing.
AB
ComponentVolume per sample
100% Isopropanol1.2 ml
1.4.

Bead Mix

Combine the components that are listed below. Prepare enough mixture for the number of samples extracted that day with allowing extra for pipetting loss.

Note
Prepare the bead mix on the day it will be used. Bead mix can be stored in the fridge or on top of ice until it is needed for up to 4 hours. Avoid freezing the mixture as it damages the magnetic beads. Be careful with using cold racks as they can cause accidental freezing of tube content.

  • Vortex the nucleic acid binding beads well to ensure they are fully resuspended.
AB
ComponentVolume per sample
Nucleic Acid Binding Beads30 μl
Lysis ENHANCER/ Proteinase K30 μl
Total volume60 μl
  • Mix well by vortexing and store appropriately until needed.
2.

Sample Processing

2.1.

Prepare the lysate in 5mL or 15mL centrifuge tubes

For each sample:

  • Set up and label 5ml Eppendorf tubes, then aliquot 2.4 ml of the lysis/binding solution (supplemented with carrier RNA and 100% Isopropanol - see step 9 and 10 in Reagent preparation) into each tube.
  • Add 1.2 ml of the samples to the tubes containing the Lysis/Binding solution. ·

Note
· When adding sample, immerse pipette tips slightly in the Lysis/Binding solution to prevent creating aerosols and rinse pipette tip.

  • Mix gently by vortexing for 30s and spin briefly to collect tube content.
3.
                              **Workflow**   **A  - Automated extraction**                                
3.1.

Setting up and running the 24 deep well plates

  1. Label two 24 deep-well plates as Plate 1 and Plate 2.
  2. Transfer the 3.6 ml of sample/lysis mixture to row A of plate 1.
  3. Add 60 μl of prepared Bead Mix to wells in row A of the 24 deep well plate 1 containing the lysed sample solution using a new tip for every addition and rinsing it with the sample solution to prevent loss of beads.

Note
A vortex mixer, shaker or rotator mixer can be used for this step or tubes can be repeatedly rotated manually. It is important to achieve sufficient mixing in this step with a visibly homogenous coloured mixture throughout to ensure efficient lysis and bead binding.
4. Centrifuge tubes briefly to collect content.

Aliquot the required reagents into the plates according to the table above:

  1. Add 900 μl of prepared Wash Solution 1 to rows B and C of Plate 1.

  2. Add 1350 μl of prepared Wash Solution 2 to rows A and B of Plate 2.

  3. Place a 12-tip comb in a 96 deep well plate in row D of Plate 1.

  4. Add 50 μl of Elution Buffer or nuclease free water to row D in Plate 2.

  5. Check to confirm that the King Fisher Duo Prime is set up with 6 tip magnet and heating block.

  6. Select the program MVRI Duo LV 1200ul and load the two plates using the A1 markings as prompted. The two plates will be on opposing sides of the turning platform in the processor.

  7. Start the program and close the front lid while the King Fisher is running.

  8. After completion of the run, a final prompt will appear to unload the plates.

  9. Press the “Check Mark”, then unload the plate containing the eluted viral RNA. Transfer the RNA

to labelled containers and store appropriately until further use.

Note
Final eluate volume varies depending on evaporation loss during warm elution step.

  1. Unload the sample plate and empty contents into beaker with disinfectant. Switch on UV light for

disinfection by selecting the program in the maintenance protocols.

Changing the magnetic head on the Kingfisher Duo Prime

  • Select and start Change Magnetic Head protocol in Maintenance protocols in the device menu. (This will position the magnet to be accessible.)
  • Unscrew and remove the screws holding the magnetic head in place and lift the magnetic head to take it out.
  • Replace the required magnetic head and tighten the screws to hold it in place.
  • Remember to also change the heating blocks as the machine doesn’t give a prompt to do so!
  • Run Check 12/6 tip protocol with a dummy test plate containing the tip comb in the required row to ensure the right positioning. Unload the test plate.
4.

Workflow B - Manual extraction

Continued from Step 2.

4.1.

Bead capture and washes:

  1. Add 60 μl of prepared bead mix to each sample tube containing the lysed sample solution using a new tip for every addition and rinsing it with the sample solution to prevent loss of beads.

Note
Use a new 200 µL tip for each addition to minimise bead loss and rinse the tip gently a few times to ensure full volume is transferred.

  1. Mix tubes thoroughly at gentle speed for 4 min to fully lyse viruses and bind RNA to beads. 
    

Note
A shaker or rotator mixer can be used for this step or tubes can be repeatedly rotated or gently vortexed manually. It is important to achieve sufficient mixing in this step with a visibly homogenous coloured mixture throughout to ensure efficient lysis and bead binding.

  1. Place processing tubes on magnet and leave for at least 3 minutes to allow for bead capture to complete when beads form a pellet against the magnet.

Note
Capture times may vary. Pellets may also be smeared on the tube’s wall as well as forming a compact body. Twisting the tube gently helps making the pellet more compact if needed.

  1. Carefully aspirate and discard supernatant without disturbing the beads.

  2. It is important to remove the lysis supernatant fully, so a brief centrifugation before collecting the remaining supernatant might be necessary.
    
  3. Remove tube from magnetic stand and place in tube rack for washing with Wash Solution 1.
    
  4. Add 300 µl Wash Solution 1, supplemented with Isopropanol to each sample and vortex at moderate speed for 30s.
    

Note
Pellets do not necessarily get fully resuspended in all samples and controls.

  1. Centrifuge briefly to collect tube content.
    
  2. Capture beads on magnet for 3-5 min or until mixture becomes clear, indicating full capture.
    
  3. Carefully aspirate and discard supernatant.

  4. Repeat steps 7-10 one more time to complete two washes with Wash Solution 1.

  5. Remove tube from magnetic stand and place in tube rack for washing with Wash Solution 2.

  6. Add 450 µl Wash Solution 2 supplemented with ethanol to each sample and vortex at moderate speed for 30s.

Note
Beads often appear granular during washing steps with Was Solution 2.
14. Centrifuge briefly to collect tube content.

  1. Capture beads on magnet for 3-5 min or until mixture becomes clear, indicating full capture.

  2. Carefully aspirate and discard supernatant.

  3. Repeat steps 13-16 to complete two washes with Wash Solution 2. It is important to completely remove the supernatant after the second wash to avoid inhibition in downstream applications.

4.2.

Drying the beads and elution:

  1. Centrifuge briefly and remove any residual solution with

fine-tipped pipette.

  1. Dry the beads by leaving the tube open for 2 minutes to

  2. Add 50 µL Elution Buffer to each sample and shake/vortex vigorously for 4 min.allow any remaining alcohol to evaporate.

Note
Mix until pellet resuspends fully; this usually happens without much problem.

  1. Centrifuge briefly to collect tube content.

  2. Capture the beads on the magnet as before and collect supernatant containing the purified RNA in labelled containers and keep on ice for immediate use or store frozen until needed.

Note
Open tubes before placing them on the magnet to avoid tube content flipping on sidewall. When this happens, remove tube from magnet and centrifuge briefly before replacing on the magnet.

Note
If beads are accidentally collected, return fluid from pipette tip and try again once supernatant is clear.

Note
Pellet might be smeared around the bottom of the tube as opposed to forming a compact pellet against the magnet. Look into the open tube from above to check if eluate is clear. Aim to collect RNA from a central position without touching the pellet.

推荐阅读

Nature Protocols
Protocols IO
Current Protocols
扫码咨询