Cleaning Protocol for Running Bacterial Samples on CytoFLEX S in Tube Mode

Jamie C Tijerina

Published: 2023-07-15 DOI: 10.17504/protocols.io.q26g7p83kgwz/v1

Abstract

The CytoFLEX S analyzer, like most other flow cytometers and cell sorters, was designed to run primarily mammalian cells. The acceptable threshold for carryover generally is <1% and this can be achieved with no special cleaning protocols between samples when running mammalian cells. However, this does not hold true for bacterial cells. In testing with K-12 E. coli cells, it was found that the carryover between samples was between 2% to nearly 4% without any special cleaning protocol between samples. This level of carryover can interfere with results.

This protocol was tested and developed to reduce carryover for K-12 E. coli or other bacterial samples to an acceptable threshold as low as .3% (<1%), to guard against contamination of the instrument which could lead to the formation of biofilms, and to prevent damage to the tube/plate switch mechanism.

Before start

Prepare 100mL of fresh 25% bleach (75mL of DI+25mL of Bleach=100mL of 25% Bleach). After dilution, do not keep the prepared stock for more than 1 month as it will lose strength over time once exposed to light and air.

Steps

Between Each Experimental Sample

1.

Manually backflush 2 times (click the backflush button and let it go through a 3-second cycle; do this 2x)

2.

Change the flow rate to “Custom” using the radio button and then move the slider to the maximum flow rate position of 240ul/minute

3.

Set the stopping rules to 30 seconds time using the radio button and typing in the time.

4.

Deselect all other stopping rules.

5.

Create a new tube in the software and give it a name that you will clearly recognize as a cleaning tube.

6.

Fill a fresh tube with between 500uL and 1mL of 25% Bleach

7.

Press the “Record” Button to run the 25% bleach for 30 seconds.

8.

Create a new tube in the software and give it a name that you will clearly recognize as a cleaning tube.

9.

Fill a fresh tube with between 500uL and 1mL of DI Water

10.

Set the stopping rules to 30 seconds time using the radio button and typing in the time

11.

Press the “Record” Button to run the DI Water for 30 seconds.

12.

Manually backflush 2 times (click the backflush button and let it go through the 3-second cycle; do this 2x)

13.

Then discard the two cleaning tubes, do not reuse the same tubes to clean between each sample.

14.

Change the flow rate back to Slow (10ul/minute) using the radio button.

15.

You are now ready to run your next sample.

16.

At the end of the experiment, you may delete all the cleaning tube files within the software by using the "Delete Multiple Tubes" function

At the End of the Experiment

17.

At the end of your experiment during the “Daily Clean,” instead of running the Beckman Coulter Flow Clean detergent (blue colored cleaning agent) for the standard 5 minutes, run it for 7 minutes by changing the number on the dialog box that appears after you begin the process using the drop-down menu to select “Daily Clean.”

18.

Place a freshly prepared tube with approximately 3mL of Flow Clean on the tube loader. Click start.

19.

When the dialog box appears indicating that the Flow Clean step has completed, remove the tube of Flow Clean from the sample loader and discard it.

20.

Install a freshly prepared tube of approximately 3mL of DI water on the tube loader. Run it for 7 minutes by changing the number on the dialog box.

21.

Do not abandon the instrument during the cleaning process. Log out of the software completely at the end of your experiment.

22.

Discard all tubes used for cleaning so that you or someone else who may use the instrument does not accidentally clean with them.

23.

Wipe down the work area, mouse, and keyboard gently with CaviWipes or 70% Ethanol Spray.