C-SOP-201: Genomic DNA Quantification using a Qubit Fluorometer

Ben Pascoe, Mihir Kekre

Published: 2023-11-06 DOI: 10.17504/protocols.io.kqdg3xb47g25/v1

Disclaimer

This protocol has been adapted from dsDNA quantification assays developed by ThermoFisher Scientific.

Abstract

In order to prepare high quality double-stranded DNA libraries for Illumina WGS, extracted genomic DNA needs to be accurately quantified. Qubit Fluorometers detect fluorescent dyes that are specific to the target of interest. These fluorescent dyes emit only when bound to specific target molecules, even at low concentrations. This provides obvious advantages over UV absorbance quantitation, which measures anything absorbing at 260 nm — DNA, RNA, protein, free nucleotides, or excess salts. Moreover, UV spectrophotometry often does not have the sensitivity to accurately measure low concentrations of DNA and RNA. The Qubit dsDNA broad-range (BR) and high-sensitivity (HS) assays are highly selective for double-stranded DNA (dsDNA) over RNA and single-stranded DNA (ssDNA). This protocol has been adapted from dsDNA quantification assays developed by ThermoFisher Scientific.Variations of this method include quantification assays from Promega:DNA quantification using the Quantus fluorometerThis is an open-access protocol distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike.

Steps

Before starting

1.

The buffer and BR/HS reagent (dye) should be stored at 4Room temperature while the standards must be stored at 4°C.

The reagent is light and humidity sensitive, it should be stored in the dark and exposed minimally during assay and sample preparation. When stored as directed, the reagent has a shelf-life of about 6 months.

2.

Prior to initiating the protocol, ensure that all active workbenches are cleaned with 80% ethanol, all relevant personal protective clothing is worn and the work area is prepared for DNA quantification according to local GLP guidelines.

Create an organised bench space by clearing away all clutter in order to maximize work efficiency.

Qubit Mastermix and Standard/Sample Preparation

3.

Set up the required number of Qubit tubes for standards and samples.

Note
i. The Qubit dsDNA assays require two standards (#1 and #2) for calibration.ii. If you have already performed a calibration on the Qubit machine for the selected assay you can use the previous calibration stored on the machine. It is recommended to perform a new calibration for every sample batch but a same-day calibration would be fine to use for multiple batches.

4.

Label the sample assay tube lids with sample IDs and the standard tubes lids as S1 and S2.

Note
i. Use only thin-wall, clear, 0.5mL PCR tubes. The recommended consumables are listed under Materials (Cat. No. Q32856).ii. Do not label the side of the tube as this could interfere with the sample reading

5.

Prepare a Qubit Mastermix for each sample and standard by diluting the Reagent in a 1:200 ratio with Buffer.

Per sample:

Qubit® dsDNA BR/HS Reagent: 1µL

Qubit® dsDNA BR/HS Buffer: 199µL

To avoid pipetting errors, it is recommended to add some overage as indicated below in Table 1.

ABCDEF
Volume per sample (µL)For 10 samples (x 11)For 20 samples (x 21)For 30 samples (x 31)For N samples
Qubit Reagent1112131N+1
Qubit Buffer199218941796169199 x (N+1)

Table 1

Note
To avoid any cross-contamination, it is recommended that the total amount of working solution required for samples and standards is removed from the working solution bottle and then added to the mastermix container instead of pipetting directly from the stock bottle to each tube.

6.

Aliquot the appropriate volume of Qubit Mastermix into each tube:

  • Standard tubes: 190µL
  • Sample tubes 198µL

Next, to the appropriate tube:

  • Add 10µL of standards (S1 and S2).
  • Add 2µL of each sample.

The final volume in each tube must be 200µL once the samples/standards are added.

Note
i. The Qubit dsDNA assays require 2 standards for every round of calibration.ii. For samples that are suspected to have low yields of DNA, the sample tube volume can be decreased to anywhere between 180-198µL in order to supplement the final volume with a greater quantity of input DNA.iii. When adding 1–2µL of sample input, use a P-2 pipette for best results.

7.

Cap each tube and then mix vigorously by vortexing for 0h 0m 5s.

Allow all tubes to incubate at Room temperature for 0h 2m 0s.

Reading Standards and Samples with the Qubit 4.0

8.

On the home screen of the Qubit, press dsDNA , then select dsDNA: High Sensitivity or dsDNA: Broad Range depending on the kit being used.

9.

INSTRUMENT CALIBRATION:

If calibration is being performed, press Read Standards to proceed. (If a calibration is not necessary, you can omit steps 8.1 to 8.3).

9.1.

Insert the tube containing Standard #1 into the sample chamber, close the lid, and press Read Standard . When reading is complete, remove the tube and place back in the rack.

9.2.

Insert the tube containing Standard #2 into the sample chamber, close the lid, then press Read Standard . When reading is complete, remove the tube and place back in the rack.

9.3.

The instrument displays the results of the calibration immediately after S1 and S2 have been read. For further information on interpretation of the results, refer to the Qubit 4.0 Fluorometer User Guide.

10.

Next, press Read Samples to proceed to the assay.

10.1.

On the Sample Volume screen, select the volume of sample added to the tube and the output concentration units (i.e. ng/μl or other).

10.2.

Insert the first sample tube into the sample chamber, close the lid, and then press Read Tube .

11.

Note
i. The software displays the results of each sample. The top value (in large font) is the concentration of the original sample. The bottom value is the dilution concentration. Record the concentration of the original sample, remove the tube and repeat steps 9.1 and 9.2 to record results for each additional sample.ii. To manage and import assay readings directly from the instrument, refer to the relevant Fluorometer User Guide in the next section.

Additional Information & Troubleshooting

12.

For more information, refer to the following links depending on the model of instrument and type of assay being used.

Quick Reference Qubit Assays

Qubit dsDNA BR Assay Guide

Qubit dsDNA HS Assay Guide

Qubit 3.0 Fluorometer User Guide

Qubit 4.0 Fluorometer User Guide

Qubit Flex Fluorometer User Guide